EXHIBITION & AWARDS  |  PRODUCTS CATALOGUE  


An ISO 9001:2008 Certified Company
Send Online Enquiry
LAMINAR FLOW BENCH HORIZONTAL
LAMINAR FLOW BENCH HORIZONTAL-STI-231 A
LAMINAR FLOW BENCH HORIZONTAL
 
 
PRODUCT APPRAISAL/APPLICATION
 
Laminar Air Flow Benches/Cabinets are Self Contained Workstation units conceptualized to provide a work Area free from any kind of Particulate Contamination. These units can be placed within the clean room to create an Ultra-Clean Environment or for providing an Individual Air Environment free from Impurities or in applications requiring protection of Specimen from possible contamination during the process or tests.
 
 
SALIENT FEATURES
 
  • Versatile Performance
  • Satin finish
  • Lower Vibration & noise level
  • Designed to provide a work Area conforming to class 100 of Revised US Federal Standards 209B
CLASSIFICATION/ CLEANLINESS LEVEL
Laminar Air Flow is designed to meet the requirements of US Federal standard 209 B providing particle free air to meet the class 100 requirements. Most useful for microbiological manipulations. For sterilized horizontal air passing through pre-filter and HEPA filter with the help of a motorised blower.
 
 
CONSTRUCTION DETAILS
  • Cabinet Construction. Whole Cabinet fabricated out of Thick Industrial Grade Commercial Board /Mild Steel Sheet or Galvanized Iron Steel Sheet duly pre treated with primers & elegantly painted with durable powder coating paint/ Stainless Steel Sheet. In models with Wooden Construction, whole cabinet is fabricated out of Industrial Grade commercial Board with all visible surfaces laminated with mica lamination. The Inner surface of the unit is coated with Air Drying epoxy paint. Working Table/surface is made of Thick Wooden Board with Stainless Steel clad on the top.
  • Front Door. The Front Door panel is made of Heavy duty Perspex sheet duly framed in Anodized Aluminium frame.
 
ILLUMINATION & ULTRA VIOLET GERMICIDAL TUBE
The unit is provided with Adequate Illumination at the work table by means of Diffused Glare free fluorescent lights fitted at the upper portion of the unit along with Ultra Violet Germicidal Tube for Sterilization inside the working area thereby ensuring high standard of cleanliness even before the commencement of actual working. Unit is supplied complete with Pre Filter, HEPA Filter, Static Pressure Manometer, Ultra Violet Germicidal Tube, Gas Cock, cord & plug. Suitable to operate on 220/230 Volts, 50 Hertz, AC supply.
 
FILTERS ASSEMBLY 
  • PRE FILTRATION: Laminar air flow is fitted with fully washable synthetic pre-filter units.
  • MAIN FILTRATION(HEPA): At the Delivery end, the unit is incorporated with Secondary High Efficiency Particulate Filters (HEPA) made of mini pleated non woven fabric.
  • FILTER DESCRIPTION: Laminar air flows are fitted with fully washable synthetic pre-filter units and secondary high efficiency perfect air filters made of pleated Non Woven Fabric. The efficiency of the filter has a rating of better than 99.99% at DOP (cold) and 99.97% at DOP (Hot). These units have the capacity to hold all suspended particles of size greater than 0.3 micron. At the delivery end, the unit is provided with HEPA Filters (High Efficiency Particulate Air Filter) comprising of Non-Woven Micro fiber Glass pleated media. The pleats are separated by corrugated Aluminium foil separators. The filter has Rubber padding for a perfect rubber to rubber sealing. Air drawn through Pre filter is made to pass through highly effective High Efficiency Particulate filter having efficiency rating as high as 99.99% with cold DOP & 99.97% with hot DOP, thus retaining all Air borne particles of size 0.3microns and larger. 
 
 
MOTOR & DYNAMICALLY BALANCED BLOWER ASSEMBLY
Incorporated with a heavy duty efficient motor which operates with minimum noise levels. Using a Dynamic Machine, the Motor & Blower Assembly is both statically and dynamically balanced ensuring minimum vibration and noise levels.
 
 
POWER REQUIREMENTS
Suitable to operate on 220/230 Volts, Single phase, 50 Hertz, AC supply.
 
 
TECHNICAL FEATURES & SPECIFICATIONS
 
 

INNER WORKING

DIMENSIONS

      2’  x  2’  x  2’ 

(600 x 600 x 600 mm)

      3’  x  2’  x  2’ 

(900 x 600 x 600 mm)

        4’  x  2’  x  2’ 

(1200 x 600 x 600 mm)

       6’  x  2’  x  2’ 

(1800 x 600 x 600 mm)

       8’  x  2’  x  2’ 

(2400 x 600 x 600 mm)

  HEPA FILTER(SIZE)

              2’  x  2’  x  6”

          3’  x  2’  x  6”

4’  x  2’  x  6”

       3’  x  2’  x  6”

      4’  x  2’  x  6”

  HEPA FILTER (No.)

                       1

                  1

1

 2

                 2     

  PRE FILTER

  FULLY WASHABLE SYNTHETIC PRE FILTERS

  HEPA FILTER

  • Glass Pleated Non Woven Fabric separated by corrugated Aluminium foil separators
  • Efficiency rating as high as 99.99% with cold DOP & 99.97% with hot DOP
  • Retaining all Air borne particles of size 0.3microns and larger

  Air Flow

  Unidirectional

  ILLUMINATION

  Diffused Glare free fluorescent light

  POWER REQUIREMENT

  220/230 Volts, 50 Hertz, AC supply

  CABINET CONSTRUCION

  COMMERCIAL GRADE WOODEN BOARD

  MILD STEEL SHEET/G.I SHEET

STAINLESS STEEL SHEET

     UPGRADES                       (Optional Accessories)

  • ULPA Filters instead of HEPA Filters
  • Timer for UV Tube
 
 

Note: Following Company’s constant endeavor to improve design &Quality the actual equipment may differ in detail, description or as depicted in this catalogue.

Testimonials


“ ”
Om Sai Ram
Socialize With Us :